CELIV Routine Commands¶
This section lists the commands specific to the CELIV routine.
Settings JSON¶
{
"output":{
"Pulse Delay":{
"Start (s)":0,
"Step (s)":1E-5,
"End (s)":1E-4
},
"Pulse Width (s)":1E-6,
"V Ramp":{
"Duration (s)":2E-4,
"Start (V)":0,
"End (V)":-1
},
"Delay (s)":0.001,
"Averages":100,
},
"device_thickness_m":2E-5,
"instrument":{
"type":"Fast",
"config":{
"range":{"value":0.05,"unit":"A"}
}
}
}
device_type
The instrument.config object changes based on the instrument.type selected:
"device_settings":{
"compliance":{"value":0.01,"unit":"A"}
},
"device_settings":{
"range":{"value":0.1,"unit":"A"}
},
Data JSON¶
{
"delay":{"value":1E-4,"unit":"s"},
"data_scheme":[
{"name":"Voltage","unit":"V"},
{"name":"Current","unit":"A"}
],
"data":[
[2.19706854771811,-2.22762695659185E-5],
[2.19749007802445,-1.38456634391332E-5],
[2.19741104109201,-1.54264020879054E-5],
[2.19712782541744,-2.10907155793393E-5],
...
],
"parameters":{
"delay":{"value":1E-4,"unit":"s"},
"t_max":{"value":7.53E-5,"unit":"s"},
"deltaJ":{"value":1.92323202267289E-5,"unit":"A"},
"J0":{"value":-2.72819419536972E-5,"unit":"A"},
"A":{"value":5000,"unit":"V/s"},
"mobility":{"value":3.15124676318999E-18,"unit":"m^2/Vs"}
}
}
Progress JSON¶
{
"routine_status":"Running",
"routine_name":"CELIV",
"progress":{
"total_delays":11,
"delays_done":4,
"averaging_pct":40.98
},
"error":null
}
Custom Commands¶
No custom commands are available for the Photo-CELIV routine
Example Command Sequence¶
{ "target": "MAIN", "command": "StartRoutine", "parameter": { "routine": "Photo-CELIV"} }
{ "target": "ROUTINE", "command": "GetTestStatus" } // repeat until routine_status == "Ready"
{ "target": "ROUTINE", "command": "ApplySettings", "parameter": { ... Settings JSON ... } }
{ "target": "ROUTINE", "command": "StartMeasurement" }
{ "target": "ROUTINE", "command": "GetTestStatus" } // repeat until routine_status != "Running"
{ "target": "ROUTINE", "command": "GetTestData" }
{ "target": "ROUTINE", "command": "CloseRoutine" }