Skip to content

Transient Photo Response Routine Commands

This section lists the commands specific to the Time-Correlated Single Photon Counting (TCSPC) routine.

Settings JSON

{}

Data JSON

Note Offset and Resolution are expressed in ns. Histogram is the counts per bin.

{"Offset":0,"Resolution":0.025,"Histogram":[6,7,11,8,3,6,9, ... ]}

Custom Commands

Command Description
GetRates Get the actual sync and input rates.
SetLaser Enable or disable the excitation laser.

GetRates

Description
Get the actual sync and input rates.

Example Request

{
  "target": "ROUTINE",
  "command": "GetRates",
  "request_id": 201
}

Example Response
Rates are expressed in counts per second

{
  "status": "OK",
  "data": {"sync_rate":10000370,"input_rate":193650,"ratio":0.0193642835215097},
  "request_id": 204
}

SetLaser

Description
Control the excitation laser

Example Request

{
  "target": "ROUTINE",
  "command": "SetLaser",
  "parameter": {"port_id":7, "laser":true},
  "request_id": 204
}

Example Response

{
  "status": "OK",
  "data": { "state":"OK" },
  "request_id": 204
}

Example Command Sequence

{ "target": "MAIN",    "command": "StartRoutine", "parameter": { "routine": "TCSPC"} }
{ "target": "ROUTINE", "command": "GetTestStatus" } // repeat until state == "Ready"
{ "target": "ROUTINE", "command": "SetLaser", "parameter": {"laser":true},}
{ "target": "ROUTINE", "command": "GetRates" } //Check that sync rate != 0 and Ratio is 1-5%
{ "target": "ROUTINE", "command": "StartMeasurement" }
{ "target": "ROUTINE", "command": "GetTestStatus" } // repeat until state == "Ready"
{ "target": "ROUTINE", "command": "GetTestData" }